Catalog Number 5C4471
Device Problem Filling Problem (1233)
Patient Problem No Patient Involvement (2645)
Event Date 11/01/2017
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation.External visual inspection was performed with no issue noted.An internal visual inspection revealed a damaged door handle, a cracked door bezel, deteriorated piston foam, a cracked piston, deteriorated membrane gasket, deteriorated pump foam pad, a damaged occluder boot, and debris in pump filter and pump tubing.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device met electrical performance specification requirements per rite testing; however failed function testing for volumetric accuracy.An accuracy confirmation test was performed and failed.The cause was deteriorated piston foams.The piston foam was to be scrapped and the device was sent for servicing.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation of a returned homechoice device, a baxter technician determined the device failed fluid volume accuracy testing.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
largo FL
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7077168
MDR Text Key93625185
Report Number1416980-2017-09520
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 12/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C4471
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2017
Date Manufacturer Received12/01/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse