• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET Back to Search Results
Catalog Number 367364
Device Problems Break (1069); Material Separation (1562); Sticking (1597)
Patient Problem Foreign Body In Patient (2687)
Event Date 02/14/2017
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Results: a sample was not returned for evaluation.A review of the device history record revealed no irregularities during the manufacture of the reported lot # 6082838.Thirty retain samples were tested for cannula pull force.The minimum force measured 28.48 lbs.With a maximum value of 33.37 lbs.Conclusion: without a sample, an absolute root cause for this incident cannot be determined as bd was not able to duplicate or confirm the customer¿s indicated failure mode.
 
Event Description
It was reported that bd vacutainer® ultratouch¿ push button blood collection set had the needle remain inside the patient while the plastic fixing device broke in half.Needle was extracted as it was sticking out a few millimeters.No injury or medical intervention reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET
Type of Device
BLOOD COLLECTION SET
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7079036
MDR Text Key94409595
Report Number1024879-2017-01128
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903673648
UDI-Public50382903673648
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K153309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 11/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2018
Device Catalogue Number367364
Device Lot Number6082838
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-