• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® PLUS PLASTIC SST TUBE; BLOOD COLLECTION TUBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® PLUS PLASTIC SST TUBE; BLOOD COLLECTION TUBE Back to Search Results
Catalog Number 367983
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/06/2016
Event Type  malfunction  
Manufacturer Narrative
Results: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.Retention samples were selected from bd inventory for evaluation, and the customer's indicated failure mode for faulty tube stoppers was not observed as all samples met specifications.A review of the manufacturing records was completed for the incident lot and no issues were identified.Conclusion: bd was unable to duplicate or confirm the customer's mode of failure.
 
Event Description
It was reported that some tube stoppers were coming off of the bd vacutainer® plus plastic sst tubes.No serious injury or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PLUS PLASTIC SST TUBE
Type of Device
BLOOD COLLECTION TUBE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7088185
MDR Text Key94805042
Report Number1024879-2017-01199
Device Sequence Number1
Product Code JKA
UDI-Device Identifier00382903679836
UDI-Public00382903679836
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K023075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number367983
Device Lot Number5356537
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/22/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-