MAKO SURGICAL CORP. TIBIAL INSERT IMPACTOR; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 160176
Device Problems Break (1069); Fracture (1260)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/08/2017
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
Pka poly impactor broke.
 
Manufacturer Narrative
An event regarding crack/ fracture involving a mako impactor was reported.The event was not confirmed.Method & results: device evaluation and results: the device was not returned for evaluation, however, a photo of the subassembly handle was provided.There is no photo of the poly impactor tip.The device has some scratches and markings throughout which appear to be consistent with usage.Medical records received and evaluation: not performed as patient factors did not contribute to the event.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies complaint history review: there has been no other event for the lot referenced.Conclusions: the event could not be confirmed nor the root cause determined because the devices were not returned for evaluation and insufficient information was provided.There is no indication of fracture based on the photo provided of the impactor handle and no photo was provided of the poly impactor tip.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.
 
Event Description
Pka poly impactor broke.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL INSERT IMPACTOR
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
timothy rice
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key7095625
MDR Text Key94963617
Report Number3005985723-2017-00610
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number160176
Device Lot Number25090609
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/26/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age58 YR
-
-