• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX INC. TRIM-IT DRILL PIN, 1.5MM X 100MM; PIN, FIXATION, SMOOTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX INC. TRIM-IT DRILL PIN, 1.5MM X 100MM; PIN, FIXATION, SMOOTH Back to Search Results
Catalog Number AR-4151DS
Device Problems Material Fragmentation (1261); Use of Device Problem (1670)
Patient Problem Foreign Body In Patient (2687)
Event Date 06/06/2017
Event Type  Injury  
Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Device history record review revealed nothing relevant to this event.At this time, it cannot be determined how the device may have caused or contributed to the incident.An evaluation of the device cannot be performed as the device was discarded by the facility.Additional information has been requested but not made available.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.This is the first complaint of this type for this part/lot combination.The potential causes of this event are being communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
 
Event Description
It was reported that the patient underwent a hammertoe repair procedure on her right fourth toe on (b)(6) 2017, using trim it pins.The pin did not perform as it should, and she had to do a second surgery on (b)(6) 2017 to remove the pin.No further details provided at time of initial report.Follow-up investigation from medical records: op report (b)(6) 2017 procedure: the patient implant was an ar-4151ds, implant delivery system, trim-it drill pin 1.5x 100 mm.Patient diagnosis which led to the (b)(6) 2017 procedure was right fourth digit hammertoe.Patient (b)(6) at time of revision.(b)(6) lbs.Height (b)(6) inches, (b)(6).Op report (b)(6) 2017 procedure: on (b)(6) 2017 patient underwent a revision surgery due to painful retained internal fixative, deep right 4th digit.Extensive dissection was needed to remove the trim-it pin.It was necessary to cut the trim-it pin at the level of the distal interphalangeal joint and then it was removed proximally and distally.The proximal portion was found to be in an odd shape and one portion was found to be in a u shape.The second small fragment was removed distally and the entire piece measured 3 cm which was felt to be the entire length of the trim-it pin.Surgeon office records dated (b)(6) 2017: patient reported non-compliance at follow up visit four days post op (b)(6) 2017 procedure.Patient reported walking without her prescribed shoe at night to the bathroom.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIM-IT DRILL PIN, 1.5MM X 100MM
Type of Device
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
ARTHREX INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath, adverse events
1370 creekside boulevard
naples, FL 34108-1945
8009337013
MDR Report Key7105364
MDR Text Key94479744
Report Number1220246-2017-00451
Device Sequence Number1
Product Code HTY
UDI-Device Identifier10888867032788
UDI-Public10888867032788
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/01/2018
Device Catalogue NumberAR-4151DS
Device Lot Number10069208
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/09/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age61 YR
Patient Weight63
-
-