Device evaluation summary: there was no damage observed to the device or the tip.There were one (1) portholes opened on the endoanchor cassette.Visual inspection confirmed there was no endoanchor loaded in the applier.The handle was opened, and no fluid or blood was observed within the handle.No corrosion was observed to the circuit board or connector pins.All wires and connectors within the handle appeared to be intact and properly connected.The battery was removed and measured 7.80v.A new battery was attached, and the unit powered up with the blue error light on, the forward light flashing and the back light on.The device was restarted in factory mode with the blue error light on, forward light flashing, back light on.An endoanchor was loaded and deployed with no abnormalities noted.If information is provided in the future, a supplemental report will be issued.
|