The subject device was not returned to olympus medical system corp.(omsc).Therefore, omsc could not evaluate the reported device.The exact cause of this issue could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.The reported event is most likely related to the operator's technique.Omsc surmised that the kinks of the outer tube occurred due to applying bending load to the outer tube when the tube was inserted into the endoscope, the subject device was unpacked from the sterile bag, or pre-use inspection was performed.The frictional resistance between the outer tube and the needle tube increased due to the kinks, and then the needle could not be retracted into the outer tube.The instruction manual of the device has already warned as follows; insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.
|