MEDTRONIC NAVIGATION, INC. (LOUISVILLE) MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM; EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION COMPACT
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 11/17/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Device serial number not available from the site.Device manufacturing date is dependent on serial number, therefore, unavailable.The axiem box was changed and the system fully functions.No parts have been returned to the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that while outside of a procedure the axiem box was working intermittently.The navigate light does not light up at times.There was no patient present when this issue was identified.No further information was provided.
 
Manufacturer Narrative
The hardware investigation of the returned axiem box found that the reported event was related to an electrical issue.Port 8 stayed amber when instrument was inserted.Eminterface reported a fault on the instrument.Tracking details showed port 8 not used.The reported event was confirmed.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Manufacturer Narrative
Correction: udi number.Additional information in: part #, manufacturer system; account name.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM
Type of Device
EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
jonathan wong
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7118779
MDR Text Key95057752
Report Number1723170-2017-05154
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 08/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION COMPACT
Device Catalogue Number9735602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/18/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-