Model Number FUSION COMPACT |
Device Problems
Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/17/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.Device serial number not available from the site.Device manufacturing date is dependent on serial number, therefore, unavailable.The axiem box was changed and the system fully functions.No parts have been returned to the manufacturer for evaluation.
|
|
Event Description
|
A medtronic representative reported that while outside of a procedure the axiem box was working intermittently.The navigate light does not light up at times.There was no patient present when this issue was identified.No further information was provided.
|
|
Manufacturer Narrative
|
The hardware investigation of the returned axiem box found that the reported event was related to an electrical issue.Port 8 stayed amber when instrument was inserted.Eminterface reported a fault on the instrument.Tracking details showed port 8 not used.The reported event was confirmed.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
|
Manufacturer Narrative
|
Correction: udi number.Additional information in: part #, manufacturer system; account name.
|
|