Catalog Number 254401006 |
Device Problem
Insufficient Information (3190)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/22/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune femoral impactor has been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.The annealed product was released on 30-jul-2014.This device is from an un-annealed batch.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.Device history lot = > null.Device history review = > null.
|
|
Event Description
|
It was reported that when the surgeon impacted the femur, the femur impactor split down the middle and broken into 2 pieces.
|
|
Manufacturer Narrative
|
(b)(4).Investigation summary = > the investigation confirmed that the impactor is cracked.Expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune femoral impactor has been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.The annealed product was released on (b)(6) 2014.This device is from an un-annealed batch.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: the investigation confirmed that the impactor is cracked.Expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune femoral impactor has been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.The annealed product was released on (b)(6) 2014.This device is from an un-annealed batch.A field safety notice was issued in (b)(6) 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.
|
|