ST. JUDE MEDICAL, COSTA RICA LTDA PRESSURE WIRE AERIS G8; TRANSDUCER, PRESSURE,CATHETER TIP Back to Search Results
Model Number C12058
Device Problems Fracture (1260); Material Fragmentation (1261)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 12/13/2017
Event Type  Injury  
Event Description
The pressurewire aeris was placed in the pda and two stents were deployed.As the pressurewire aeris was being removed from the anatomy the coil came off in the left pda and it took 3 hours to snare the coils.There was no part of the device left in the anatomy.There was no resistance noted when removing the wire.But the physician thinks the coils might have been trapped in the distal architecture of the vessel.The procedure was completed and the patient was fine when he went to recovery.The only patient information given was that the patient was male.Further information was not provided due to hospital privacy policy.
 
Manufacturer Narrative
Product evaluation: the reported event of "the tip coil came off" was not confirmed.The results of the investigation concluded that the tip coil damage was unable to be determined; however, the hydrophilic coated distal tube, microcables, and corewire had been fractured.The fractured guidewire section was not returned.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the tip coil damage remains unknown; however, the hydrophilic coated distal tube, microcables, and tip coil damage is consistent with forcible contact during use.The pressurewire instructions for use (ifu) states that excessive manipulation of the pressurewire when the sensor element or pressurewire tip is located in sharp bend may cause damage or tip fracture.The pressurewire instructions for use (ifu) states that the pressurewire is a delicate instrument and should be handled carefully.Bending or excessive force during removal from the packaging coil may damage the guidewire.The pressurewire instructions for use (ifu) states that torqueing the pressurewire against resistance or repeated attempts to cross a total vessel occlusion may cause damage and/or fracture, which may lead to a portion of pressurewire separating from the tip.The pressurewire instructions for use (ifu) states that the user should advance or withdraw the pressurewire slowly and never push, withdraw or torque the pressurewire if it meets resistance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE WIRE AERIS G8
Type of Device
TRANSDUCER, PRESSURE,CATHETER TIP
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7121921
MDR Text Key94988936
Report Number3008452825-2017-00338
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131452
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Model NumberC12058
Device Catalogue NumberC12058
Device Lot Number6013872
Other Device ID Number05414734055772
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-