MAUDE Adverse Event Report: BD (SUZHOU) BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM; INTERVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD (SUZHOU) BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM; INTERVASCULAR CATHETER Back to Search Results
Catalog Number 383406
Device Problems Out-Of-Box Failure (2311); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/21/2017
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported by the nurse that the bd intima-ii¿ closed iv catheter system had brownish-red foreign matter inside the un-opened packaging.Found before use.No serious injury or medical intervention noted.
 
Manufacturer Narrative
Results - bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for foreign matter with the incident lot was observed.The foreign matter was seen outside of the liquid flow path.It looks to be lubricating oil.A review of the manufacturing records was completed for the incident lot and no issues were identified.Conclusion - the most probably root cause is the foreign matter was on the production line which resulted in the defect.However, since the foreign matter was found outside the fluid pathway this is not a significant issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM
Type of Device
INTERVASCULAR CATHETER
Manufacturer (Section D)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer (Section G)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7123336
MDR Text Key95477204
Report Number3006948883-2017-00166
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K143610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date06/23/2020
Device Catalogue Number383406
Device Lot Number7139386
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-