Catalog Number 383406 |
Device Problems
Out-Of-Box Failure (2311); Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/21/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported by the nurse that the bd intima-ii¿ closed iv catheter system had brownish-red foreign matter inside the un-opened packaging.Found before use.No serious injury or medical intervention noted.
|
|
Manufacturer Narrative
|
Results - bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for foreign matter with the incident lot was observed.The foreign matter was seen outside of the liquid flow path.It looks to be lubricating oil.A review of the manufacturing records was completed for the incident lot and no issues were identified.Conclusion - the most probably root cause is the foreign matter was on the production line which resulted in the defect.However, since the foreign matter was found outside the fluid pathway this is not a significant issue.
|
|