MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Air Leak (1008)
Patient Problems Stroke/CVA (1770); Seizures (2063); Numbness (2415)
Event Date 03/13/2017
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryoablation procedure, as the balloon catheter was inserted in the sheath, air ingress occurred.The case was completed with cryo.The patient was then diagnosed with a stroke.The patient's hospitalization was extended.No further patient complications have been reported as a result of this event.Additional information indicated that a medication was administered to treat the epilepsy.The patient experienced numbness of the left hand, and a computed tomography (ct) scan was performed, which confirmed the stroke.It was also reported that the sheath was used with the balloon catheter when the air ingress occurred.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Event summary: multiple clinical issues were encountered during the procedure.(air ingress, seizure and stroke).The flexcath sheath 4fc12/ 59822 was not returned for investigation.In conclusion, this is a case related to clinical issues (air ingress, seizure and stroke) encountered during procedure.The sheath was not returned for investigation.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7124424
MDR Text Key95175899
Report Number3002648230-2017-00684
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/16/2018
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number59822
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/16/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age58 YR
-
-