• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH 2.0MM ASNIS MICRO, CANNULATED SCREWDRIVER, 2.0MM, AO COUPLING; ELBOW JOINT RADIAL (HEMI-ELBOW) POLYMER PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH 2.0MM ASNIS MICRO, CANNULATED SCREWDRIVER, 2.0MM, AO COUPLING; ELBOW JOINT RADIAL (HEMI-ELBOW) POLYMER PROSTHESIS Back to Search Results
Catalog Number 45-20001
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/04/2017
Event Type  malfunction  
Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event Description
Two of item 45-20001 were broken in the tip during insertion of the screw.
 
Manufacturer Narrative
The reported event that 2.0mm asnis micro, cannulated screwdriver, 2.0mm, ao coupling was alleged of issue (breakage during surgery) could be confirmed, since the device was returned for evaluation and matches the reported failure mode.The screwdriver was received broken at the tip.In general, it looks quite used.The edges of the remaining helix are slightly twisted and deformed with the device having scratch marks on its surface.The breakage surface appears to be dull and the patterns are consistent with over-torqueing of the instrument.The screwdriver was most likely twisted under high loads and eventually broke.Based on the investigation, the case could be classified as user-related due to over-torque.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
 
Event Description
Two of item 45-20001 were broken in the tip during insertion of the screw.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM ASNIS MICRO, CANNULATED SCREWDRIVER, 2.0MM, AO COUPLING
Type of Device
ELBOW JOINT RADIAL (HEMI-ELBOW) POLYMER PROSTHESIS
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
Manufacturer (Section G)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
Manufacturer Contact
rose haas
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key7124530
MDR Text Key95195722
Report Number0008031020-2017-00701
Device Sequence Number1
Product Code KWI
Combination Product (y/n)N
PMA/PMN Number
K023604
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number45-20001
Device Lot NumberAH02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-