The reported event that 2.0mm asnis micro, cannulated screwdriver, 2.0mm, ao coupling was alleged of issue (breakage during surgery) could be confirmed, since the device was returned for evaluation and matches the reported failure mode.The screwdriver was received broken at the tip.In general, it looks quite used.The edges of the remaining helix are slightly twisted and deformed with the device having scratch marks on its surface.The breakage surface appears to be dull and the patterns are consistent with over-torqueing of the instrument.The screwdriver was most likely twisted under high loads and eventually broke.Based on the investigation, the case could be classified as user-related due to over-torque.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
|