Model Number BI-700-00027-120
Device Problems Computer Software Problem (1112); Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 11/22/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative reported that reloading the software on image acquisition system (ias) resolved the issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts were replaced.No parts have been received by the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that while outside of procedure, the generator and motion status of the image acquisition system (ias) would not load and the progress bar displayed zero progress.Rebooting the system multiple times did not resolve the issue.Representative attempted connecting to image acquisition system (ias) via remote but an error message was received.There was no patient present at the time of the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7126694
MDR Text Key95193158
Report Number1723170-2017-05206
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Is the Reporter a Health Professional? No
Date Manufacturer Received11/22/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/06/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-