Catalog Number 00515048201 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/06/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
This complaint is being reported by zimmer biomet as (b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during surgery, the battery pack was "fevered" (hot) when it was opened.Therefore, the surgeon used an alternate product for the surgery.No adverse events have been reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
Current leakage problem is for the "battery pack leak".Review of the device history record identified no relevant deviations or anomalies.Product examination found that the battery pack had failed because a battery had leaked.This caused corrosion of the battery contacts.This complaint is confirmed.Reviewing the complaints for the lot number z000008541 shows 1 complaints for this part and lot number.The root cause of the reported event could not be specifically determined with the provided information.However, the cause of this battery failure could not be determined.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
|
|