• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HIP KIT; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HIP KIT; LAVAGE, JET Back to Search Results
Catalog Number 00515048201
Device Problem Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/06/2017
Event Type  malfunction  
Manufacturer Narrative
This complaint is being reported by zimmer biomet as (b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during surgery, the battery pack was "fevered" (hot) when it was opened.Therefore, the surgeon used an alternate product for the surgery.No adverse events have been reported as a result of this malfunction.
 
Manufacturer Narrative
Current leakage problem is for the "battery pack leak".Review of the device history record identified no relevant deviations or anomalies.Product examination found that the battery pack had failed because a battery had leaked.This caused corrosion of the battery contacts.This complaint is confirmed.Reviewing the complaints for the lot number z000008541 shows 1 complaints for this part and lot number.The root cause of the reported event could not be specifically determined with the provided information.However, the cause of this battery failure could not be determined.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIP KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key7128777
MDR Text Key96007448
Report Number0001526350-2017-00995
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2020
Device Catalogue Number00515048201
Device Lot NumberZ000008541
Other Device ID Number00889024375192
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received02/19/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-