• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 ACETABULAR SCREW; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 ACETABULAR SCREW; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source:foreign.The event occurred in (b)(6).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Not returned to manufacturer.
 
Event Description
It was reported the screw could not be fully threaded.The surgery was completed with another device, causing a surgical delay of 15 minutes.No further information has been made available.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
The following report is being submitted to relay additional information received the complaint can be confirmed based on the returned complaint sample.Visual inspection found the screw head to be damaged.A portion of the head's rim material has been pulled away and detached from the head.The hex feature has also been deformed, indentations can be seen on the inside of the hex.The device history records were reviewed and no related deviations/ anomalies were identified that affect the reported event.A definite root cause cannot be identified if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 ACETABULAR SCREW
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7128790
MDR Text Key95752215
Report Number0001825034-2017-11026
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number010000999
Device Lot Number6083438
Other Device ID Number(01) 0 0880304 52747 8
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A