Model Number 165816 |
Device Problem
Physical Resistance (2578)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
|
Event Description
|
It was reported that upon deflation and removal, the catheter balloon appeared to be asymmetrical.The complainant alleged that there was some resistance of the catheter when trying to remove, but the catheter was removed without medical intervention.
|
|
Manufacturer Narrative
|
The reported event was confirmed as cause unknown.Received 2 portions of a silicone catheter for evaluation.Also, received a photo showing a catheter with an asymmetrical balloon.Per visual inspection, no defects were found and it was observed that the shaft and inflation arm with valve are cut.Further inspection noted that there was a cut at the distal end of the shaft to 2 ¿ 5/8¿ (2.625¿) of the silicone catheter tip and inflation arm section with valve measured 1 - 1/2" (1.5¿).No other defects were observed.However, per visual inspection of the picture received, balloon asymmetrical was noted.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "recommended inflation capacities 3cc balloon: use 5ml sterile water 5cc balloon: use 10ml sterile water 30cc balloon: use 35ml sterile water do not exceed recommended capacities.Visually inspect the product for any imperfections or surface deterioration prior to use." (b)(4).
|
|
Event Description
|
It was reported that upon deflation and removal, the catheter balloon appeared to be asymmetrical.The complainant alleged that there was some resistance of the catheter when trying to remove, but the catheter was removed without medical intervention.
|
|
Search Alerts/Recalls
|