Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Concomitant medical products: 110017124, g7 finned bm 4 hole shell 54f, 3589847.Report source, foreign ¿ events occurred in (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the liner would not fix to the shell.Attempts have been made and additional information on the reported event is unavailable.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Liner was returned for review.Visual inspection of the part showed nicks, gouges, and dings on the liner face, outside surface, and scallops.Also, frayed poly & few indentation marks identified on the outer surface of the liner dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no trends were identified.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|