Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical products: 110017124, g7 finned bm 4 hole shell 54f, 3589847.Report source, foreign ¿ events occurred in (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the liner would not fix to the shell.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Liner was returned for review.Visual inspection of the part showed nicks, gouges, and dings on the liner face, outside surface, and scallops.Also, frayed poly & few indentation marks identified on the outer surface of the liner dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no trends were identified.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL E1 LINER 32MM F
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7130388
MDR Text Key95797787
Report Number0001825034-2017-11100
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date06/11/2022
Device Model NumberN/A
Device Catalogue Number010000850
Device Lot Number6074803
Other Device ID Number(01) 0 0880304526365
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A