Model Number 0570-0313
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/01/2017
Event Type  malfunction  
Manufacturer Narrative
Upon review of the device history for the serial number (b)(4), it was determined that the device was manufactured on (b)(6) 2011 and the one (1) year warranty period has expired.Due to the age of the device and the fact there are no repairs available for the video baton the customer elected to replace the glidescope avl video baton 3-4.The video baton was not returned to verathon for evaluation, therefore the cause could not be conclusively determined.However, it is likely that the age of the device, six (6) years, caused or contributed to the event.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event Description
The customer reported that during a patient, using a glidescope avl video baton 3-4, the video baton was not recognized by the video monitor and would not display an image.Reportedly second video baton was used with the same video monitor to complete the procedure, however the amount of time needed to prepare the second video baton was not reported.No harm to patient or user was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE AVL VIDEO BATON 3-4
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key7135251
MDR Text Key95913036
Report Number9615393-2017-00231
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0570-0313
Device Catalogue Number0570-0313
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/04/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/22/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-