Model Number 295025-001
Device Problems Detachment Of Device Component (1104); Mechanical Problem (1384)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 12/12/2017
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the battery was in the battery charger and not in use in the patient's freedom driver at the time of the customer-reported issue.The freedom onboard battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the "prongs" on the bottom of the patient's freedom onboard battery pulled outward when removed from the battery charger.
 
Manufacturer Narrative
Visual inspection of the battery did not reveal any evidence of damage to the connector bodies or contacts.Investigation efforts to confirm the unseated / loose connector issue reported by the customer were unsuccessful; the condition could not be duplicated or reproduced.The freedom onboard battery was tested in accordance with the current evaluation process, where it successfully passed all sections.The onboard battery performed as intended with no evidence of a device malfunction.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM ONBOARD BATTERY
Type of Device
BATTERY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
kerri hensley
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key7138742
MDR Text Key96118879
Report Number3003761017-2017-00257
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003084
UDI-Public(01)00858000003084
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 10/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number295025-001
Device Catalogue Number295025-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/12/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/22/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age27 YR
-
-