Catalog Number 7068398
Device Problem Metal Shedding Debris (1804)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/25/2017
Event Type  malfunction  
Manufacturer Narrative
This part is not approved for use in the united states; however a like device catalog # 5440030, 510k # k102555 and (b)(4) was cleared in the united states.Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.
 
Event Description
Pre-operative diagnosis :l1 burst fracture levels :t11-l2 it was reported that during surgery, when the tightening of the set screw was done without a sequential reducer,the thread of the set screw was found to be stripped.The right l2 set screw was removed once, and checked.There was a delay of less than 60 minutes in overall procedural time.No fragments of the set screw were remaining in the patient.No patient complications were reported.
 
Manufacturer Narrative
Product analysis: set screw appears to be the screw that was misaligned and caused the splaying and thread damage.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
stacie ziemba
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key7142207
MDR Text Key95745704
Report Number1030489-2017-02560
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/17/2023
Device Catalogue Number7068398
Device Lot NumberH5194905
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-