Product complaint # (b)(4).The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Evaluation: a representative sample was returned for analysis.This product code is double armed and required four strands per packet.Upon visual inspection, no defects were found on the package.The sample was opened the swage and attachment area were as expected and the needles/sutures were found correctly placed.The sutures were dispensed without problems and examined along the strand, no defects and/or damaged were found.The sutures were measured in diameter and the met the finished goods requirements in minimum individual and average for this product code.In addition, the samples were tested by needle pull and met the finished goods requirement for this product code.Per the condition of the representative sample, no assembly error, incorrect component or performance pull off suture needle were found on the sample and the tested needles/suture combinations met the finished goods requirements.
|