OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-XP180N
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2017
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.Omsc reviewed the manufacturing history of this device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of routine microbiological testing by the user facility, the instrument channel of the subject device tested positive for microbes (7cfu/100ml).The type of microbes are unknown.This device had been reprocessed using an olympus automated endoscope reprocessor model etd3 (not available in the usa) with peracetic acid.The user facility reported that this device was reprocessed according to the instruction manual.There was no report of patient infection associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to olympus medical systems corp.(omsc) but was returned to olympus (b)(4).(b)(4) sent this device to a third party laboratory for additional microbiological testing.As a result of additional microbiological testing, the following results were obtained.Therefore, it cleared the french guideline.Total channel: 2 cfu/100ml (micrococcaceae).Instrument channel: no microbe.Suction channel: no microbe.Air/water channel : 5 cfu/100ml (micrococcaceae).The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Manufacturer Narrative
This supplemental report is submitting to correct "device product code".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
Manufacturer Contact
katsuaki morita
2951 ishikawa-cho
hachioji-shi, tokyo-to 
426425177
MDR Report Key7145285
MDR Text Key96219243
Report Number8010047-2017-02070
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK100584
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/24/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-XP180N
Other Device ID Number04953170339097
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/13/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/25/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-