Catalog Number NS-9008 |
Device Problem
Programming Issue (3014)
|
Patient Problem
No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).It has been communicated that the device is not available for evaluation.Without the device it is not possible for codman to conduct a proper investigation.Since a lot number has been provided a review of the manufacturing records will be reviewed.We anticipate that the evaluation will reveal that the device conformed to specifications prior to release.If anything otherwise is found then a follow up report will be filed.If at some point the device does become available, this complaint will be re-opened, evaluated and a follow up report will be filed.Trends will be monitored for this and similar complaints.At the present time this complaint is considered closed.Device not available.
|
|
Event Description
|
The valve was implanted to the patient via lp-shunt about a month ago.Doi and initial setting were unknown.It was reported that the surgeon tried to change the pressure setting but it was not able to be changed.The revision surgery was performed on (b)(6) 2017.The patient¿s initial is (b)(6), female but her age is unknown.There was no surgical delay and no adverse consequence to the patient.No further information was provided by the hospital.The product will not be returned to your site.
|
|
Manufacturer Narrative
|
It was initially reported that the device would be returned for evaluation.However, multiple attempts to obtain the sample were unsuccessful.This report has been updated to reflect the corrected information.Complaint sample was not returned to codman and no product or lot number information has been provided; therefore, an evaluation of the device could not be performed and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined.If the complaint sample becomes available, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this or similar complaints.At present, we consider this complaint to be closed.
|
|
Manufacturer Narrative
|
It was incorrectly reported in additional mfr narrative that the device had not been made available for evaluation.The device has been returned and will be evaluated.Upon completion of the investigation a follow-up report will be submitted.
|
|
Manufacturer Narrative
|
(b)(4).Upon completion of the investigation it was noted that the images were taken of the ¿as received¿ valve.The position of the cam when valve was received was 130 mmh2o.The valve was visually inspected: no defects were noted.The valve was hydrated.The valve was tested for programming with programmer 82-3126 with serial (b)(4), the valve passed the test.The valve was flushed, the valve passed the test no occlusion was noted.The valve was leak tested.Leaked from the needle holes in the needle chamber.The valve was reflux tested.The valve passed the test.The siphon guard was tested.The siphon guard passed the test.The valve was dried.The siphon guard was removed.Review of the history device records for the valve, product code ns9008, with lot ctgbjf, conformed to the specifications when released to stock on the 12th june 2015.No root cause could be determined; as no problem was noted with the valve.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
|
|