• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. UNIFY ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. UNIFY ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3361-40C
Device Problem Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/22/2017
Event Type  Injury  
Event Description
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.There were no adverse consequences to the patient prior, during and post procedure.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7152183
MDR Text Key95963225
Report Number2938836-2017-35469
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2017
Device Model NumberCD3361-40C
Device Lot NumberS000003722
Other Device ID Number05414734508223
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0116-2017
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age52 YR
-
-