Results: one sample unit was received for evaluation by our quality engineer team.Upon examination, the needle was observed to be partially retracted and damage was found on the grip and at the connection of the grip to the barrel.The needle was pushed and repositioned to the outer position, the button was depressed, and the retraction was unsuccessful.Although the review of the dhrs review are required for mdr¿s per cpr 071, a review could not be performed as the lot number was not provided.Conclusion: the root cause of the defect was determined to have occurred during the manufacturing process.Within the production zones, the plug probe and load barrel stations have the ability to become misaligned and produce the type of damage observed.Proper personnel have been notified of this incident.
|