• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG SOLUTIONS S.A. MAHURKAR; CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG SOLUTIONS S.A. MAHURKAR; CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED Back to Search Results
Model Number 8888345629
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/14/2017
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, there was difficulty during the passage of the catheter where it folds, collapses and there was absence of flow impacting the functioning.There was no harm to the patient.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted no abnormalities.Pmv performed functional testing which included submerging the catheter into a water bath.The end was clamped, and a syringe was used to inject air to observe leakage.No air bubble was present.Both extension was tested with acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Type of Device
CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
Manufacturer (Section G)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
Manufacturer Contact
lisa hernandez
15 hampshire st.
mansfield, MA 02048
2034925563
MDR Report Key7156758
MDR Text Key96219693
Report Number3009211636-2018-00002
Device Sequence Number1
Product Code NIE
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K102605
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/09/2018
Device Model Number8888345629
Device Catalogue Number8888345629
Device Lot Number312919X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/24/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
-
-