Model Number 989605414121
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Return of the suspect transducer is anticipated.Evaluation results will be included in a follow-up report upon its return and investigation completion.
 
Event Description
A customer reported intermittent articulation issues with their x7-2t model transducer.There was no patient injury associated with this event.
 
Manufacturer Narrative
Evaluation of the transducer could not duplicate the articulation issue as described by the customer.Investigation of the device noted a hole in the bending neck sheath, scuff mark on the lens and debris on the handle switch.Additional functional testing of the device confirmed the mechanical movement of this device was functioning as designed and no articulation impediment was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
ULTRASOUND
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
paul corrigan
22100 bothell everett highway
bothell, WA 98021
4254877000
MDR Report Key7159029
MDR Text Key96229020
Report Number3019216-2018-00001
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K030455
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 10/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number989605414121
Device Lot NumberB1DMPY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-