Model Number BI-700-00027-120
Device Problems Computer Software Problem (1112); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(6).A medtronic representative inspected the imaging system onsite and found that the left side of the bellows cover was not properly aligned in the track.The bellows cover was placed into the correct track and the issue resolved.No parts were replaced or returned to the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that while in a spinal fusion procedure, the imaging system requested a motion calibration upon boot up.The system was in the operating room (or) and the site was able to run it and then continue with the procedure.There was a reported delay to the procedure of less than 1 hour due to this issue and no impact on the patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
stephanie riley
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7161533
MDR Text Key96464592
Report Number1723170-2018-00030
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 01/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/06/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/04/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age70 YR
Patient Weight106
-
-