C.R. BARD, INC. (COVINGTON) -1018233 LUBRI-SIL® I.C. COMPLETE CARE® URINE METER FOLEY TRAY; URINE DRAINAGE BAG (CONOD) Back to Search Results
Catalog Number 903114A
Device Problem No Flow (2991)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/08/2017
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that urine in the meter would not flow into the bag.The urine would flow out of the catheter into the meter successfully, but when the meter became filled, the urine would not flow into the bag.The user believed that there was an issue with the air filter.
 
Manufacturer Narrative
The reported issue was unconfirmed.The device was returned and visually inspected.Observed 3 needles were punctured on the urine meter air.Evaluation verified no abnormalities observed on the returned sample.Evaluation was conducted by introducing water into the urine meter to verify the flow rate of water out from the drainage bag.As result, the water was out smoothly without any restriction.Further, evaluation removed all the needles and sealed the air vent on the urine meter by using a clear tape prior to introducing water into the urine meter and drainage bag.Water was introduced again to verify the flow rate and water flowed out smoothly without any restriction.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use state the following: (b)(4).
 
Event Description
It was reported that urine in the meter would not flow into the bag.The urine would flow out of the catheter into the meter successfully, however when the meter became filled, the urine would not flow into the bag.The user believed that there was an issue with the air filter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUBRI-SIL® I.C. COMPLETE CARE® URINE METER FOLEY TRAY
Type of Device
URINE DRAINAGE BAG (CONOD)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7161572
MDR Text Key96473373
Report Number1018233-2018-00017
Device Sequence Number1
Product Code MJC
Combination Product (y/n)N
PMA/PMN Number
K984136
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number903114A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2018
Date Manufacturer Received03/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-