Catalog Number 004411100
Device Problems Overheating of Device (1437); Temperature Problem (3022)
Patient Problems Burn(s) (1757); No Patient Involvement (2645)
Event Date 12/21/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device involved in this complaint has not been received by the manufacturer at the time of this report.The investigation into this complaint is in progress.
 
Event Description
Customer complaint alleges "handle malfunction - bulk (bulb) staying on and over heating handle even when blade is not engaged".There was no report of patient involvement.Reporter states user was burned.There was no report of necessary medical intervention.
 
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event Description
Customer complaint alleges "handle malfunction - bulk (bulb) staying on and over heating handle even when blade is not engaged".There was no report of patient involvement.Reporter states user was burned.There was no report of necessary medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUSCH GREENSPEC FO HANDLE MEDIU
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
TRUPHATEK INTERNATIONAL LTD.
14 benny gaon street
p.o. 8051
netanya 42504 43
IS   4250443
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key7164184
MDR Text Key96378533
Report Number8030121-2018-00002
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number004411100
Device Lot Number170401
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-