Device 2 of 2.Reference mfr.Report: 1627487-2018-00015.It was reported the patient experienced random unpleasant sensations on his right side.The issue was reproduced by the physician upon palpation of the extension to ipg site.Subsequently, surgical intervention was undertaken to explant and replace the ipg and extension.During the procedure, the physician reported broken sutures at the ipg site which resulted in the ipg moving freely in the pocket.In addition, the extension appeared to be bent.The issue has resolved following the procedure.
|