SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED IMMULITE 2000 HCG Back to Search Results
Model Number IMMULITE 2000 HCG
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/06/2017
Event Type  malfunction  
Manufacturer Narrative
A siemens customer service engineer (cse) was dispatched to the customer site.After analyzing the instrument, the cse performed system decontamination with sodium hydroxide solution.The cse then reviewed the mechanical positions of probes, dual resolution dilutors and verified hydraulics couplings.The siemens technical application specialist performed repeatability testing on the patient sample.The customer stated that the patient had a breast cancer and any positive hcg result is considered important for the treatment of the patient.As per immulite 2000 hcg instructions for use, the intended use of hcg is "for in vitro diagnostic use with the immulite® 2000 systems analyzers - for the quantitative measurement of human chorionic gonadotropin (hcg) in serum, and for strictly qualitative determinations in urine, as an aid in the detection of pregnancy." therefore, hcg was being run by the customer for an off-label use.The cause of the discordant, falsely elevated hcg results on one patient sample is unknown.Siemens is investigating the issue.Mdr 2432235-2018-00008 was filed for the same event.Mdr 2432235-2018-00009 was filed for the event occurred on (b)(6) 2017.
 
Event Description
The customer obtained discordant, falsely elevated human chorionic gonadotropin (hcg) results on one patient sample on an immulite 2000 xpi instrument, while using reagent lot 386 and 387.The results from the immulite 2000 xpi instrument were reported to the physician(s).The sample was repeated on an alternate platform, resulting lower and matching the clinical history of the patient.The corrected result from the alternate platform was reported to the physician(s).There are no reports of patient intervention or adverse health consequence due to the discordant, falsely elevated hcg results.
 
Manufacturer Narrative
The initial mdr 2432235-2018-00012 was filed on 04-jan-2018.Additional information (04-jan-2018): the sample id in question was (b)(6).Sections have been updated with this information.The customer uses human chorionic gonadotropin (hcg) for tumor marker detection.Quality control was within acceptable range.No sample is available for in-house testing.Mdrs 2432235-2018-00008_s1, 2432235-2018-00009_s1, 2432235-2018-00040, 2432235-2018-00041, and 2432235-2018-00042 were filed for the same issue.
 
Manufacturer Narrative
The initial mdr 2432235-2018-00012 was filed on 04-jan-2018.The first supplemental mdr 2432235-2018-00012_s1 was filed on 30-jan-2018.Additional information (07-feb-2018): a siemens headquarters support center (hsc) specialist reviewed the event data.The human chorionic gonadotropin (hcg) results were reproducible on immulite 2000 xpi instrument across multiple draws for the sample in question.Quality control and other patient samples were not affected.The hsc specialist stated the issue only affected this patient sample.A list of medication taken by the patient is not available.Hcg reagent has been formulated to minimize the risk of interference.However, potential interactions between rare sera and test components can occur.For diagnostic purposes, the results obtained from this assay should always be used in combination with the clinical examination, patient medical history, and other finding.Based on the information provided, there is no known device issue.The device is performing within manufacturing specifications.No further evaluation of device is required.Mdrs 2432235-2018-00008_s2, 2432235-2018-00009_s2, 2432235-2018-00040_s1, 2432235-2018-00041_s1, and 2432235-2018-00042_s1 were filed for the same issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMMULITE 2000 HCG
Type of Device
IMMULITE 2000 HCG
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED
glyn rhonwy
llanberis
gwynedd, LL55 4EL
UK  LL55 4EL
MDR Report Key7165720
MDR Text Key96726694
Report Number2432235-2018-00012
Device Sequence Number1
Product Code DHA
UDI-Device Identifier00630414961132
UDI-Public00630414961132
Combination Product (y/n)N
PMA/PMN Number
K990222
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup,Followup
Report Date 02/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIMMULITE 2000 HCG
Device Catalogue NumberL2KCG2
Device Lot Number387
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/07/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-