• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEVICOR MEDICAL PRODUCTS, INC. MICROMARK BIOPSY SITE MARKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEVICOR MEDICAL PRODUCTS, INC. MICROMARK BIOPSY SITE MARKER Back to Search Results
Model Number C1535
Device Problem Material Fragmentation (1261)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2017
Event Type  malfunction  
Manufacturer Narrative
The c1535 micromark® ii tissue marker is indicated for use to attach to soft breast tissue at the surgical site during an open surgical breast biopsy or a percutaneous breast biopsy to radiographically mark the location of the biopsy procedure.The micromark ii tissue marker is a sterile, single patient use device comprised of a small stainless steel marker and a disposable introducer and applier.The introducer consists of a flexible tube, a distal ramp, and a lateral port.The applier consists of a flexible shaft, a deployment mechanism, and a handle.The marker is located at the distal end of the applier.The device has not been returned to the manufacturer for evaluation which prevents a full investigation and analysis of the root cause at this time.However, based on the instructions for use, the following could cause or contribute to reported event.Significant force to the applier handle when inserting device into probe for deployment.Removing the device independently from the probe when significant resistance is encountered.Twisting or kinking of the applier shaft.The treating physician indicated no fragments of the device remained in the patient.However, as the occurrence has the potential to result in patient consequence with the need for surgical intervention or other form of treatment, should fragments have been transferred to the patient and pursuant to 21 cfr 803, we are submitting this medwatch report.
 
Event Description
The (b)(6) affiliate reported that the doctor deployed the clip successfully.When the doctor pulled the c1535, the flexible introducer was broken and dropped.The doctor told us that there are not break of introducer in patient body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROMARK BIOPSY SITE MARKER
Type of Device
BIOPSY SITE MARKER
Manufacturer (Section D)
DEVICOR MEDICAL PRODUCTS, INC.
300 e-business way
fifth floor
cincinnati OH 45241
Manufacturer (Section G)
DEVICOR MEDICAL PRODUCTS DE MEXICO
sor juana ines de la cruz
#20152 4-b, parque industrial
tijuana, baja california 22440
MX   22440
Manufacturer Contact
shawna rose
300 e-business way
fifth floor
cincinnati, OH 45241
5138649178
MDR Report Key7168820
MDR Text Key97591689
Report Number3008492462-2018-00002
Device Sequence Number1
Product Code NEU
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K013413
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 12/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/18/2022
Device Model NumberC1535
Device Catalogue NumberC1535
Device Lot NumberF11625301D
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-