Model Number 397002-001 |
Device Problems
Failure of Device to Self-Test (2937); Noise, Audible (3273)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/21/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver has been returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the companion 2 driver did not pass the system check and made an unusual noise.
|
|
Manufacturer Narrative
|
Review of the patient data file confirmed the three customer-reported system checks but the failures were not able to be reproduced and the driver passed all functional testing and passed three additional system checks.The customer-reported unusual noise was confirmed during investigation testing and the root cause was determined to be a malfunction of the cooling fan.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|