Catalog Number BI70002000
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Electrical /Electronic Property Problem (1198); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 12/11/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative inspected the imaging system onsite and found the motions not initializing.The problem was tracked to a loss of 24 volt dc to the brakes in the power conversion module.The power conversion module was replaced and the issue resolved.A full imaging system check-out was completed following part replacement and all tests passed.Full system functionality was confirmed and the system was returned to service.The suspect part has been returned to the manufacturer for evaluation, however, no results are available at this time.
 
Event Description
A medtronic representative reported that while outside of a procedure, the imaging system was not initializing.The representative attempted to trouble shoot the mobile view station (mvs) desktop, but the system was down.
 
Manufacturer Narrative
The enclosure power conversion was returned to the manufacturer for analysis.Analysis found that there was a loss of 24 volt dc to brakes in the power conversion module.There was a transistor found to be shorted.Analysis found that the reported event was related to an electrical issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
pa chia vue
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7169195
MDR Text Key96608580
Report Number1723170-2018-00074
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-