Model Number N/A |
Device Problems
Difficult to Insert (1316); Material Deformation (2976)
|
Patient Problem
No Code Available (3191)
|
Event Date 12/12/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(6).The reported event is confirmed as the photographs showed damage of the threads and the threads were also worn out.The taper adaptor body showed wear and missing component.No products were returned; therefore, the visual and dimensional inspections were not performed.Device history record (dhr) was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required.The potential field age of the product is approximately 2 years 2 months with unknown frequncy of usage.A definitive root cause cannot be determined with the information provided.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that during hip arthroplasty revision, the extractor threads were damaged and stripped and were unable to be assembled with the mating implant.A 60 minute delay was noted, as the well fixed stem had to be removed to successfully complete the procedure.Attempts have been made and additional information on the reported event is unavailable.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned device confirmed that the removal threaded shaft was fractured in two pieces and the distal threads of the handle assembly were damage and shows signs of wear and tear.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|
|
|