Model Number N/A
Device Problems Difficult to Insert (1316); Material Deformation (2976)
Patient Problem No Code Available (3191)
Event Date 12/12/2017
Event Type  Injury  
Manufacturer Narrative
(b)(6).The reported event is confirmed as the photographs showed damage of the threads and the threads were also worn out.The taper adaptor body showed wear and missing component.No products were returned; therefore, the visual and dimensional inspections were not performed.Device history record (dhr) was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required.The potential field age of the product is approximately 2 years 2 months with unknown frequncy of usage.A definitive root cause cannot be determined with the information provided.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that during hip arthroplasty revision, the extractor threads were damaged and stripped and were unable to be assembled with the mating implant.A 60 minute delay was noted, as the well fixed stem had to be removed to successfully complete the procedure.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned device confirmed that the removal threaded shaft was fractured in two pieces and the distal threads of the handle assembly were damage and shows signs of wear and tear.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPER ADAPTOR REMOVAL TOOL
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7169970
MDR Text Key96581702
Report Number0001825034-2017-11499
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
PMA/PMN Number
PK082996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 09/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number110005232
Device Lot Number679270
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2018
Date Manufacturer Received08/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-