Catalog Number BI70002000 |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/12/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was unavailable from the site.A medtronic representative inspected the imaging system onsite and found that there was intermittent failure to acquire fluoro images while the tube is in the lateral position.The representative traced the problem to the mfov.The detector moving while trying to acquire an image.The mfov assembly was replaced and the issue resolved.A full imaging system check-out was completed following part replacement and all tests passed.Full system functionality was confirmed and the system was returned to service.The suspect part has not been returned to the manufacturer for evaluation.
|
|
Event Description
|
A medtronic representative reported that while in a case, the imaging system went into standalone mode.There was a reported delay to the procedure of less than 1 hour due to this issue.No additional information was provided.
|
|
Manufacturer Narrative
|
The positioner for the imaging system was returned to the manufacturer for analysis.The positioner was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
|
|
Manufacturer Narrative
|
Additional information: patient demographics provided.There was a reported delay to the procedure of ten minutes due to this issue.
|
|
Manufacturer Narrative
|
The rotor control box was returned to the manufacturer for analysis.The rotor control box was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
|
|
Search Alerts/Recalls
|