The device was not returned to the manufacturer for physical evaluation, therefore the failure mode cannot be confirmed.However, an investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, the device history record review confirmed the labeling, material, and process controls were within specification.
|
An integra territory manager reported that on (b)(6) 2017, the needle tip portion of the puncture needle with irrigation bent 90 degrees when trying to insert it into the gallbladder by the surgeon during a laparoscopic cholecystectomy.Upon removing the instrument from the patient, the needle tip portion broke off completely and fell onto the floor.There was no patient injury or complications noted.However, there was a delay in surgery since the clinical team was trying to search for the broken tip.
|