CINCINNATI SUB-ZERO, LLC. HEMOTHERM 400 CE; CARDIOPULMONARY BYPASS TEMPERATURE CONTROLLER Back to Search Results
Model Number 400CE
Device Problem Device Issue (2379)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/17/2017
Event Type  malfunction  
Manufacturer Narrative
Csz medical technical support received an email stating the device went down during a procedure.The device was switched out and the procedure finished successfully.No patient harm or injury occurred.(b)(4), a csz 3rd party technician, found the power board and membrane needed to be replaced.The (b)(4) technician replaced the power board and membrane.The device was tested and functions as designed.
 
Event Description
Csz received an email stating the device went down during a procedure.The device was switched out and the procedure finished successfully.No patient harm or injury occurred.This report was filed in our complaint handling system as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOTHERM 400 CE
Type of Device
CARDIOPULMONARY BYPASS TEMPERATURE CONTROLLER
Manufacturer (Section D)
CINCINNATI SUB-ZERO, LLC.
12011 mosteller road
cincinnati OH 45241
Manufacturer Contact
christina miracle
12011 mosteller road
cincinnati, OH 45241
MDR Report Key7174963
MDR Text Key97490788
Report Number1516825-2018-00001
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122813
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number400CE
Device Catalogue Number86022
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-