It was reported that ventricle perforation and death occurred.Transcatheter aortic valve implantation (tavi) was performed using a 26mm diameter non-bsc valve.An amplatz wire was used.When the valve was deployed for about 1/3, the blood pressure of the patient decreased and left ventricular perforation was observed through echo.While preparing the pcps (percutaneous cardiopulmonary assist device), the patient was observed to have cardiac arrest.Then, pcps was connected and surgical procedure was performed.At that time, the valve was implanted.Since the heart torn over a wide range, artificial heart-lung device was used and suturing was performed using pericardial patch.After closing the wound, decannulation was not performed and the procedure was completed with the pcps being attached as it was.On the same day, the patient died due to circulatory failure.The physician considered that one of the causes of the left ventricular perforation was that the guidewire was pushed more than the expected.
|