MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Catalog Number BI70002000
Device Problems Computer Software Problem (1112); Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 12/13/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.No parts have returned to the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that while outside of a procedure, the imaging system was not starting up.The pendent shows "system booting, please wait".After 1h the same message displayed.The site restarted the system and it boots up normally.There was no patient present when this issue was identified.
 
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The imaging system passed the system checkout and was found to be fully functional.The software investigation found that a probable cause was unable to be determined without further information since the behavior could not be replicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
amos jarrette
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208903200
MDR Report Key7177424
MDR Text Key96817673
Report Number1723170-2018-00147
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Is the Reporter a Health Professional? No
Date Manufacturer Received12/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-