Model Number DIMENSION® RXL MAX WITH HM |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/15/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer reported discordant tacrolimus results on patient samples.Calibrations and quality controls (qc) were within acceptable range.A siemens customer service engineer was dispatched to the customer site.Upon the regional support center (rsc) specialist's recommendations, the following troubleshooting steps were performed: system check and diagnostic test for photometer alignment were run, lamp calibration values were verified, reagent 1 (r1), reagent 2 (r2) and sample probes alignments to vessel points were checked, and ultrasonic performance for the probes were verified.The customer ran precision testing, which was acceptable.Upon the rsc specialist's recommendation, the cse replaced the vessel feeder.The cse found that the customer mixes sample by hand.A siemens headquarter support center (hsc) specialist reviewed the instrument data and determined that it was consistent with the mixing issues.The hsc specialist stated, it is important that the samples are well mixed.Mixing by hand is uncontrolled and inefficient.Use of a rocker or a spiral mixer is the recommended procedure for mixing whole blood samples that are used for cell counts or for chemistry determinations.Based on the instrument data, the hsc specialist determined that the cause of the discordant, falsely elevated tacrolimus result on three patient samples was due to the sample related issue.The instrument is performing according to the specifications.No further evaluation of this device is required.
|
|
Event Description
|
Discordant, falsely elevated tacrolimus results were obtained on three patient samples on a dimension rxl max with hm instrument.The initial results were not reported to the physician(s).New samples were obtained from the patient and tested on a dimension exl located at another laboratory, resulting lower.The results from the dimension exl instrument were reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated tacrolimus results.
|
|
Manufacturer Narrative
|
The initial mdr 2517506-2018-00018 was filed on january 11, 2018.Additional information (01/23/2018): a siemens technical applications specialist (tas) guided the customer on how to properly work with the tacrolimus method, including running adequate sample volume, using a roller to mix samples instead of mixing by hand, using an automatic pipettor for aspiration of qc and patient samples, and processing sample tubes as soon as they are removed from the mixer.
|
|
Search Alerts/Recalls
|
|