Catalog Number 420739
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
A customer from (b)(6) reported a false susceptible flucytosine result for a candida norvengensis external quality control sample (eeq biologie prospective) in association with the vitek® 2 ast-ys08 test kit.The ast-ys08 card results were: fluconazol mic = 16 intermediate, flucytosine mic = 4 susceptible.The customer reported that the expected results were resistant for both fluconazol and flucytosine.A biomérieux internal investigation will be initiated.
 
Manufacturer Narrative
The correct pma/510(k) number is k133952.
 
Manufacturer Narrative
This report was initially submitted in response to a customer from (b)(6) reporting a false susceptible flucytosine result for a candida norvengensis external quality control sample (eeq biologie prospective) in association with the vitek® 2 ast-ys08 test kit.An internal biomérieux investigation was performed that included testing for both fluconazole (flu) and flucytosine (fct) results with a strain of candida norvegensis on vitek® 2 ast-ys08 cards.This strain is from a biologie prospective survey (mycology 2017- sample 3a).Id testing : identification was confirmed to candida norvegensis on vitek 2 yst id card (lot 2430356103).Ast testing: · reference method to determine the intended result for fluconazole and flucytosine: broth microdilution, method used to develop the formularies "flu02n" / "fct02n" (in ast-ys08 card).· flu mic = 16 mg/l intermediate · fct mic = 8 mg/l intermediate interpretation according to vitek 2 breakpoints in v7.01 : fda 2009 breakpoints for candida / fluconazole s /=64.Fda 2012 breakpoints for candida / flucytosine : s /=32.Note: for antifungals, essential agreement between vitek® 2 card and reference method mics is +/- 2 doubling dilutions.· products & system incriminated: -ast -ys08 on vitek 2 v7.01 three (3) vitek 2 ast-ys08 lots were tested from sabouraud dextrose agar (sda).Vitek 2 gave the following results on the three (3) lots tested: **flu mic = 8 mg/l susceptible **fct mic = 4 mg/l susceptible the vitek 2 results are within essential agreement with the reference mic within +/- 1 doubling dilution.Conclusion: -the customer results were duplicated on the ast-ys08 cards (mic flu= 8 mg/l s and fct mic=4 mg/l s).-the vitek 2 ast-ys08 cards perform as intended.The vitek 2 values are within essential agreement with the reference results obtained in bmd.-the reference mic for both fluconazole and flucytosine are in the middle of the breakpoints, so the strain can be susceptible or resistant within +/- 2 doubling dilutions.-->borderline strain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITEK® 2 AST-YS08 TEST KIT
Type of Device
VITEK® 2 AST-YS08 TEST KIT
Manufacturer (Section D)
BIOMERIEUX, INC
595 anglum road
st. louis MO 63042
Manufacturer (Section G)
BIOMERIEUX, INC
595 anglum road
st. louis MO 63042
Manufacturer Contact
ellen weltmer
595 anglum road
st. louis, MO 63042
3147317301
MDR Report Key7182865
MDR Text Key97757257
Report Number1950204-2018-00020
Device Sequence Number1
Product Code NGZ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K133817
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/19/2018
Device Catalogue Number420739
Device Lot Number2880297203
Other Device ID Number03573026535353
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/20/2017
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-