Model Number 37800 |
Device Problems
Electro-Static Discharge (2149); Electromagnetic Compatibility Problem (2927)
|
Patient Problems
Discomfort (2330); Electric Shock (2554)
|
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The patient reported that they were working with an oxygen refill compressor on (b)(6) 2018 and felt "a little discomfort, like a jolt".They noted that it was nothing alarming and they were not horrible pain.The patient also reported that they felt ¿real discomfort¿ during a thunder storm last summer where they felt the jolt/discomfort sensation.The patent had their device checked when they had the issues.The patient noted that the jolting/discomfort sensation had occurred many times in the past.No further complications were reported/anticipated.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a consumer.It was reported that the cause of the patient¿s jolt/discomfort was due to an oxygen refill compressor that was new at a family members¿ home in (b)(6) 2018, and lightning/thunderstorms in the spring/summer of 2017.No further complications were reported/anticipated.
|
|