• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER TMT TRABECULAR METAL POSTERIOR FEMORAL AUGMENT BLOCK; TM POSTERIOR AUGMENT BLOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER TMT TRABECULAR METAL POSTERIOR FEMORAL AUGMENT BLOCK; TM POSTERIOR AUGMENT BLOCK Back to Search Results
Model Number N/A
Device Problems Loose or Intermittent Connection (1371); Metal Shedding Debris (1804); Device Dislodged or Dislocated (2923)
Patient Problems Pain (1994); Foreign Body In Patient (2687)
Event Date 11/01/2017
Event Type  Injury  
Event Description
It was reported that the patient underwent initial surgery on (b)(6) 2015.Almost 2 years later patient complained that his knee still had pain subsequently patient was revised on (b)(6) 2017 due to pain, wear, implant disassociation, loosening, metallosis, osteolysis.Tm augment is the only tm design controlled part.
 
Manufacturer Narrative
Patient is a (b)(6) male, medium build, no height, weight or activity level was reported.Patient has a history of 5 surgeries on his left knee spanning a 13 year period which includes the primary surgery in 2004 through the fourth revision surgery in 2017.Complaint alleges pain in the left knee, implant wear, implant disassociation, loosening, metallosis and osteolysis.The patient was revised on (b)(6) 2017.An engineering investigation was conducted by product development.The evidence while limited and not overwhelmingly conclusive, suggests that the suspected damage was primarily concentrated on the medial side of the femoral component where both a 5mm distal femoral augment block and a 5mm tm posterior femoral augment block were attached.Damage included a raised lip surrounding the internal thread on the femoral component where the distal augment screw could have pulled out from the internal threads as well as damage to the visible edge of the medial tm posterior femoral augment block.In addition, there appeared to be wear marks on the medial side of the femoral component where the 5mm distal femoral augment block was once attached.This implies that at least the nexgen rhk/lcck femoral augment screw which attached the 5mm distal femoral augment block loosened at some point in time which allowed relative motion between the two components.However, the independent radiologist who assessed the images reported that there was no evidence of wear or metallic particles seen on the films.Loosening of the nexgen rhk/lcck femoral augment screw was likely a result of failing to use the required augment torque wrench, as specified in the surgical technique and package insert, which ensures proper tightening.Failing to use the appropriate instrumentation was reported in a follow-up email from the sales rep.The 10mm distal femoral augment was also reported as being separated from the femoral augment and although not unreasonable to expect given that the same instrumentation was used, the limited evidence provided in this complaint does not support this.This investigation by product development is considered closed at this time.Should additional information become available to further this investigation, this report shall be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRABECULAR METAL POSTERIOR FEMORAL AUGMENT BLOCK
Type of Device
TM POSTERIOR AUGMENT BLOCK
Manufacturer (Section D)
ZIMMER TMT
10 pomeroy road
parsippany NJ 07054
MDR Report Key7184014
MDR Text Key97052335
Report Number3005751028-2018-00005
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
PK040487
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2019
Device Model NumberN/A
Device Catalogue Number00549003601
Device Lot Number62633359
Other Device ID NumberN/A
Date Manufacturer Received01/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-