The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.The stent was returned with the microcathether used in procedure and was deployed and jammed inside the catheter hub.Visual and microscopic inspection of the device noted the stent delivery wire (sdw) was deformed, broken, and protruding from the introducer sheath.Damage was also noted to the introducer sheath.The stent could not be advanced for inspection and functional testing could not be performed due to the condition of the returned device.Information reported by the facility indicated that during the transfer of the stent into the hub of the microcatheter, the stent became stuck.Also, information available indicated that both the stent and microcatheter were inspected and confirmed to be in good condition prior to use.Based on the information available and investigation results, it is likely the stent deformed during the transfer into the hub of the microcatheter, resulting in the sdw break and subsequently limiting its performance.Therefore, an assignable cause of operational context was assigned to this investigation.
|