• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - ENGLEWOOD 3000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - ENGLEWOOD 3000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938741
Device Problems Fluid/Blood Leak (1250); Cut In Material (2454)
Patient Problem No Patient Involvement (2645)
Event Date 12/05/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 3000ml exactamix bag leaked from a slit at the upper right corner seal of the bag.This occurred after compounding but before using with a patient.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.Microscopic inspection revealed a small tear at the upper right weld seam of the bag.Functional testing was then performed by filling the device with water, and revealed a leak from the tear.The reported condition was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE - ENGLEWOOD
englewood CO
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7188220
MDR Text Key97539683
Report Number1416980-2017-10320
Device Sequence Number1
Product Code LHI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938741
Device Lot Number1229213
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/16/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-