Model Number 1000096
Device Problems Out-Of-Box Failure (2311); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 12/23/2017
Event Type  malfunction  
Manufacturer Narrative
The device has been received for evaluation; however, device evaluation is not yet complete.A supplemental report will be filed upon completion of the evaluation.
 
Event Description
It was reported that the wake button had stopped functioning.During troubleshooting with tandem technical support, the customer confirmed that the pump turned on when plugged into a power source.There was no patient involvement, as the pump was not being used on a customer at the time of the reported event.Reportedly, the customer had an alternate pump available for alternate insulin therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM G5 SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
TANDEM DIABETES CARE
11045 roselle street
san diego CA 92121
Manufacturer Contact
julia kensick
11045 roselle st.
san diego, CA 92121
8583666900
MDR Report Key7189710
MDR Text Key97586612
Report Number3013756811-2018-00367
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 01/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1000096
Device Catalogue Number1000885
Other Device ID Number(01)00853052007257
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received01/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
-
-