• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CPI - DEL CARIBE ENDOTAK RELIANCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CPI - DEL CARIBE ENDOTAK RELIANCE Back to Search Results
Model Number 0295
Device Problems Over-Sensing (1438); Ambient Noise Problem (2877); Device Contamination With Biological Material (2908)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/27/2017
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that the patient was implanted with this cardiac resynchronization therapy defibrillator (crt-d) and right ventricular (rv) lead.After the pocket was closed, additional testing was performed and the scrub technician applied pressure on the pocket.Oversensing was found.An x-ray taken was reviewed and the terminal pin of the lead appeared seated appropriately in the header.The pocket was then reopened and the lead was reconnected.There was some fluid that came out of the seal plug, therefore air was suspected.No oversensing was noted after reconnecting the lead to the header.The pocket was reclosed and no further issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7191715
MDR Text Key97295502
Report Number2124215-2017-21483
Device Sequence Number1
Product Code NVY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910073/S077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 10/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/05/2019
Device Model Number0295
Other Device ID NumberRELIANCE 4-SITE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received10/31/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age75 YR
-
-